YL9111 Binary Pump
The cut surface of tubing should be cut at right angle without dust, tube should not be contracted, and
middle inner diameter shall not be blocked.
In order to cut stainless steel tubing, tubing cutter should be used, plastic tubing cutter or shaving cutter
should be used for teflon and similar material of tubing, and the surface should be clean and have no
crumbling.
[Fig. 5] Fitting for 1/16” OD tubing
Inlet tubing of pump connects with degassing module using 1/4-28 fitting or with solvent bottle directly,
the outlet tubing connects with injector. The fitting for injector is different depending on the injector type.