YL9160 PDA Detector
through the new tube to a waste reservoir. This will clean any particulates or oil that may be residing in
the tube that could clog the heat exchanger or contaminate the sample cell of the detector.
2-4. Connection of remote cable
YL9160 PDA detector has connection terminals for remote input/output, and marker input/output.
Note :
Please do not connect wires between cables at your discretion. If you want to connect with
the other instrument, please check input/output information and confirm with the terminal
configuration of YL9160 PDA detector.
[Fig. 2] Pin configuration of each terminal
[Fig. 3] and [Fig. 4] are the diagram of remote and the other terminal input/output. In between YL9100
series modules, connect directly and confirm the configuration with the other modules.