YL9120 UV/Vis Detector
[Fig. 3] Diagram of Remote and Marker output
[Remote operation]
START-IN
: Operate instrument, and start running of gradient program.
If you connect it with autosampler or external valve, automatic running is available.
START-OUT : If the signal input on the START-IN terminal, the signal pulse output through this port. It can
be used for synchronization of remote start with the other instrument.
MARK-IN
: To control event program or operate additional operation.
MARK-OUT
: To control time event program output.
READY-IN : To change error state and stop operation if there is a input.
READY-OUT : When YL9120 UV/Vis Detector is not ready state because of running status, output
error signal is indicated if there is a leak.
2-5. Connection of communication cable
YL9120 UV/Vis detector provides TCP/IP internet protocol as a standard. The IP address of YL9100 series
is 10.10.10.20, if DIP SW settings on the rear side are On position. If you change the IP address using
control software, the DIP SW has to be set OFF.
6
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......