YL9101 Vacuum Degasser
Y
L9101
Chapter 4. Maintenance
4-1. Filtration of solvent
Be sure to filter solvent by <5
μ
m when specially use buffering solution. Keep filtered solvent in the bottle,
which is cleared from minute particles. When use filtered solvent, which kept more than 1 week, please
filter again before use.
4-2. Caution
1) Do not pressurize the vacuum degasser module. More than 120 kPa pressure will damage degassing
membrane inside. When you put solvent bottles 2 m higher than degasser, the pressure is
approximately 120 kPa. So place the bottles less than 2 m lower.
2) Be sure to prevent from any contamination when you connect or disconnect tubes. Entering fine dust
will reduce degassing efficiency greatly.
3) If you want to increase degassing efficiency, make parallel connection of input port to output port. Do
not connect it by serial way.
4) Can not use YL9101 Vacuum Degasser for output of solvent delivery pump.
4-3. Troubleshooting
1) Error and malfunctioning
Errors of YL9101 Vacuum Degasser are occurred when ……
a) Not approach normal vacuum state 15 minutes after power on.
b) Rapid pressure drops by solvent leak during normal operation.
When above errors occurred, degasser will turn red LED. Open above solvent reservoir and check if
vacuum lines are installed correctly. If you cannot find any thing wrong, please contact our representative
or us. In order to re-start, turn off and wait for more than 5 seconds and turn on.
2) Continuous air bubbling
If air bubbles are continuously happened after turning YL9101 Vacuum Degasser on more than 1 hour,
check if input and output port and valve fitting are completed connected. Do not tighten the screw. Fitting
and joint part can be damaged. If the problem is not solved, please contact our representative or us.
5
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......