YL9160 PDA Detector
Chapter 4. Maintenance
4-1. D2 lamp
The Deuterium Lamp (D
2
) is covered from 190 to 600 nm of wavelength. The replacement time is either in
case you use the lamp more than 2000 hours or in case the light intensity is a half (50%) of the original
intensity when you installed the lamp at first. An using time of D2 lamp is counted in the system indicating
the total hours of operation after lamp on.
Check the lamp intensity as follows:
1) Power on the unit if it is not already on. Wait for a period of approximately 10 minutes.
2) Set the wavelength to 254 nm by YL-Clarity software
3) Select the Reference Light Intensity.
4) If the displayed value is less than a half of the original intensity, the lamp should be replaced to
new one. Generally, you have to consider the lamp exchange when the light intensity of reference
energy is less than 50 nA.
4-1-1. How to remove the D
2
lamp
1) Make sure that the power cord is disconnected from the rear panel of the detector.
Y
L9160
2) Unscrew the screws and remove the lamp assembly on the right front panel.
CAUTION
; UV light can damage eyes and skin. Always disconnect the power cord before
working in the vicinity of the lamp. The D
2
lamp gets quite hot. Care must be taken while handling it
to prevent from burning. Always allow the lamp to cool before removing it.
3) Disconnect the UV lamp from the detector by gently pulling it straight back toward you. DO NOT
twist the connector while pulling.
4) Unscrew the two thumbscrews holding the lamp mount in place, and pull the lamp mount straight
back
13
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......