YL9111 Binary Pump
Chapter 3. Installation
3-1. Inspection and site preparation
YL9111 Binary Pump is delivered along with the following parts when being shipped. Before opening
transportation package, perform inspection for trace of shock or mistake, and if there is abnormality, do not
open the contents and inform this company of it. And, if contents are opened, perform inspection for
existence of shock in the contents and contact with this company when trace of shock is found.
The Gradient Pump is a delicate instrument, so use original box and buffer material as far as possible
when re-packing it to transport instrument. If it is impossible to use original box; wrap pump with several
layers of buffer material, and fill the bottom, top and all other sides of pump with buffer material in order to
make pump endure shock or vibration during transportation.
Standard configuration of YL9111 Binary pump
1) Main body of instrument
2) Power cord and fuse
3) Tubing
60cm,
A. Bio Narrow Bore : ID 0.01” , OD 1/16” PEEK
B. Analytical , Semi-prep
: ID 0.02”, OD 1/16” SUS316
4) Installation
kit
5) Manual
Site requirement of YL9111 Binary pump
1) Room with 20
℃
temperature with variation ± 5
℃
with and 60% humidity
2) Where no direct and straight sunlight
3) Plain floor without carpet
4) Where having spare space of 20cm or more
5) Where there are 10% or less voltage change, no frequency change and 100
Ω
or less grounding
point
6) Where there is no generation of corrosive gas and ventilation is well done
7) Where stable power of 110 or 220V is supplied,
±
10%
8) Where not receive electromagnetic induction from large transformer, high frequency heater, UPS,
etc.
9) Within 2500 m above sea level(storage within 4600m)
8
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......