YL9130 Column Compartment
2-4. Connection of remote cable
YL9130 column compartment has connection terminals for remote input/output, external solenoid valve,
and marker input/output. The remote cable from the injector(manual or autosampler) has to be connected
on the Remote In terminal on the rear side of YL9130 column compartment to collect data at the moment
of injection.
Notice ! :
Please do not connect wires between cables at your discretion. If you want to connect
with the other instrument, please check input/output information and confirm with YL9130 column
compartment terminal configuration.
Y
L9130
[Fig. 4] Pin configuration of each terminal
<Fig 6> and <Fig 7> are the diagram of remote and the other terminal input/output. In between YL9100
series modules, connect directly and confirm the configuration with the other modules.
[Fig. 5] Diagram of Remote and Marker input
5