YL9130 Column Compartment
Chapter 4. Maintenance
YL9130 shows the error message, please refer to the following,
Error
Cause
How to fix
High Temp Limit! Temperature is higher than Max. limit
Increase max. temperature limit
Decrease operating temperature
Solvent Leak!
Solvent leaks
Remove the leaks on the leak sensor and
find the leak position
The followings are the remedy for problem.
Problem
Cause
How to fix
Actual temp. is much higher than
Setup temp.
Electronic problem Ask to manufacturer
Temperature does not increase
Heater
Instrument stops operation
1) Over heat
2) Fuse
3) Electronic problem
1) Turn off for 10minutes
2) Replace fuse
3) Ask to manufacturer