YL9111 Binary Pump
Liquid is leaked
or pressure
reduces when
cleaning port is
not used
-High pressure seal was worn.
-Plunger was worn.
-Replace high pressure seal.
-Replace plunger.
Liquid is leaked
from inside of
instrument.
-Fitting in instrument was loosened.
-Damper was damaged.
-Low pressure seal was worn.
-Tighten fitting in instrument.
-Replace.
-Replace low pressure seal.
ump is not
operated after
input of power.
-Power voltage is unstable or low.
-Strong induction voltage is
generated in the surroundings.
-Use stable, proper DC power source.
Y
L9111
-Close induction power source in
surroundings, or install instrument away
from power source.
Noise is too
high
-Load is caused to piston carrier.
-Timing belt is loose.
-Motor is in defect.
-Supply lubricant in carrier body.
-Reduce clearance above belt by
adjusting guide location of timing belt.
-Inspect connection status of motor cable,
and replace motor if noise is severe only in
high pressure.
31
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......