YL9170 RI Detector
4-1. Adjusting the optical system
To perform the adjustment the sample and reference chamber of the flow cell have to be purged with
distilled water. When both chambers contain the same liquid theoretically the same light intensity should
reach the sample and reference side of the light sensor. In this case the optical balance should be zero.
The adjustment of the optical system is performed manually. During this procedure the position of the light
beam on the photo sensor is changed by adjusting the mirror.
During the procedure you can follow the changes by observing the change of the detector signal or more
detailed by observing the difference and sum voltage.
To adjust the optical system follow the steps listed below.
-
Switch to purge mode, make sure the red purge LED is on.
-
Flush both chambers of the flow cell with clean mobile phase for several minutes.
-
Switch back to measuring mode, make sure the red purge LED is off.
-
Put the hex-wrench into one of the opened hole on the rear side.
-
Adjust the screws until the optical balance reaches a value of 0
±
20 %.
There are two adjustment screws for the mirror. Keep in mind that you might have to turn the
other screw. Anyway you should turn the screws not more that 1/2 or 3/4 turns.
10
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......