YL9110 Quaternary Pump
3-4. Connection of remote cable
YL9110 quaternary pump has connection terminals for remote input/output, external solenoid valve, and
marker input/output. The remote cable from the injector(manual or autosampler) has to be connected on
the Remote In terminal on the rear side of YL9110 quaternary pump to collect data at the moment of
injection.
[Fig. 8] Connection of remote cable between YL9110 pump and injector
Notice ! :
Please do not connect wires between cables at your discretion. If you want to connect
with the other instrument, please check input/output information and confirm with YL9110
quaternary pump terminal configuration.
[Fig. 9] Pin configuration of each terminal