EX1200A Series Introduction
25
S
YSTEM
S
PEED
Max. reading rate*
Max. internal trigger rate*
Max. Ext. trig. rate to memory*
2000/s (see
DC Voltage Reading Rates
tables below)
2000/s
2000/s
* Auto-zero off, trigger and sample delay = 0, 0.01 PLC
C
ONNECTOR
I
NFORMATION
T
YPE
5 banana jacks
Connector pin locations and signals can be found in Figure 4-1.
DMM
R
EMOVAL
/R
EPLACEMENT
VTI recommends annual calibration of the EX1200-2165 and EX1200-2365 DMMs to assure that
the instrument maintains the tightest specifications. To eliminate the need to return an entire
mainframe for calibration to VTI, the DMMs can be removed from mainframe. The mainframes
can still be used once the DMM has been removed.
Part Number
Description
Compatible Mainframe
70-0462-100
Final Assy, EX1200-2165 DMM 1U 6.5 Digit
EX1206A
70-0462-200
Final Assy, EX1200-2365 DMM 3U 6.5 Digit
EX1208A
The following sections describe how to remove the DMM from the two mainframes.
NOTE
As these devices are static sensitive, proper ESD precautions should be exercised when removing
and handling the DMM.
EX1200-2165 Removal
Removal Procedure
1)
Remove the three screws shown in using the #2 Phillips screwdriver. Retain these screws for
re-installation.
2)
Use the handle on the front panel of the DMM to slowly remove the DMM from the chassis.
Note that once half of the DMM has been removed, remove the connector at J6 on the input
board.
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......