HUNT
R
ESISTANCE
C
URRENT
R
ANGE
(ACI
&
DCI)
1 mA (DCI only)
10 mA & 100 mA
1 A & 3 A
68.1
Ω
4.1
Ω
0.1
Ω
DCV,
DCI,
2W-R,
AND
4W-R
R
EADING
R
ATE
*
I
NTEGRATION
T
IME
(PLC)
M
AX
.
R
EADINGS
/S
ECOND
[60
H
Z
(50
H
Z
)]
D
IGITS
N
ORMAL
-M
ODE
N
OISE
R
EJECTION
(AC
NMR)
DC
C
OMMON
M
ODE
R
EJECTION
(DC
CMR)
10
PLC
5.9 (4.9)
6 ½
60 dB
140 dB
1
PLC
59 (49)
5 ½
60 dB
140 dB
0.1
PLC
599 (499)
4 ½
0 dB
80 dB
0.01
PLC
2000 (1666)
3 ½
0 dB
80 dB
* Auto-zero off, trigger and sample delay = 0
AC
V
OLTAGE
–
6
½
D
IGITS
–
1
Y
EAR
A
CCURACY
R
ANGE
R
ESOLUTION
F
REQUENCY
R
ANGE
T
C
AL
5 °C
0
°C
TO
+55 °C
%
R
EADING
%
R
ANGE
<
30%
R
ANGE
>
30%
R
ANGE
100 mV
0.1 µV
3 Hz - 5 Hz
1.0000%
0.0400%
(0.1% Rdng + 40ppm Rng)/°C
>5 Hz - 20 Hz
0.3500%
0.0400%
(0.04% Rdng + 40ppm Rng)/°C
>20 Hz - 45 Hz
0.1000%
0.0500%
375
µ
V
2.25 mV
>45 Hz - 20 kHz
0.1000%
0.0500%
195
µ
V
0.81 mV
>20 kHz - 50 kHz
0.6000%
0.0800%
282
µ
V
2.82 mV
>50 kHz - 100 kHz
0.6000%
0.0800%
282
µ
V
2.82 mV
>100 kHz - 300 kHz
5.0000%
0.5000%
(0.2% Rdng + 0.02% Rng)/°C
1 V
10 V
100 V
1 µV
10 µV
100 µV
3 Hz - 5 Hz
1.0000%
0.0300%
(0.1% Rdng + 40ppm Rng)/°C
>5 Hz - 20 Hz
0.3500%
0.0300%
(0.1% Rdng + 40ppm Rng)/°C
>20 Hz - 45 Hz
0.0600%
0.0300%
0.225% Range
2.25% Range
>45 Hz - 20 kHz
0.0600%
0.0300%
0.081% Range
0.81% Range
>20 kHz - 50 kHz
0.1200%
0.0500%
0.282% Range
2.82% Range
>50 kHz - 100 kHz
0.6000%
0.0800%
0.282% Range
2.82% Range
>100 kHz - 300 kHz
5.0000%
0.5000%
(0.2% Rdng + 0.02% Rng)/°C
300 V
100 µV
3 Hz - 5 Hz
1.0000%
0.0900%
(0.1% Rdng + 40ppm Rng)/°C
>5 Hz - 20 Hz
0.3500%
0.0900%
(0.1% Rdng + 40ppm Rng)/°C
>20 Hz - 45 Hz
0.0600%
0.0900%
3.60 V
3.60 V
>45 Hz - 20 kHz
0.0600%
0.0900%
0.99 V
0.99 V
>20 kHz - 50 kHz
0.1200%
0.1500%
3.75 V
3.75 V
>50 kHz - 100 kHz
0.6000%
0.2400%
3.75 V
3.75 V
>100 kHz - 300 kHz
5.0000%
1.5000%
(0.2% Rdng + 0.02% Rng)/°C
Specifications are for sine wave input > 5% of range for 100 mV
–
100 V ranges, >15% for 300 V range, with Low AC filter selected.
Volt-hertz product is limited to 1.5x10
7
V•Hz.
Over-range: 20% beyond range, except 300 V range.
Input Impedance: 1
MΩ
2% in parallel with < 100 pF.
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......