www.vtiinstruments.com
EX1200A Web Page Operation
135
LXI Module to Module Parameters
o
LXI Domain
: an 8-bit number, 0 to 255, that indicates the domain the EX1200A is using
for LAN events. All events sent by this device will include this data. Likewise, the device
will only accept events that include the same LXI Domain data.
Wired Trigger Parameters
o
Wired-OR Bias
: indicates whether the Wired-OR bias for each Trigger Bus channel
(LXI0 through LXI7) is enabled or disabled.
LXI
I
DENTIFICATION
This status page provides the user information about the EX1200A mainframe as well as all the
plug-in modules that are installed. Firmware, driver, and network information can also be
collected from this page. Note that the page name adjusts to identify the model of the mainframe
(
as seen in “
EX1266 LXI Identification
F
IGURE
7-19:
LXI
I
DENTIFICATION
P
AGE