VTI Instruments Corp.
148
LAN/GPIB Gateway
PC
LAN Port
Ethernet
Cable
GPIB
Cable
GPIB
Cable
Daisy
Chain
GPIB
Cable
Daisy
Chain
Interconnect Cable
Function Generator
8.5 Digit DMM
Amplifier
Calibrator
LAN Port
GPIB Port
GPIB Port
GPIB Port
GPIB Port
Ethernet
Switch
Ethernet
Cable
LAN Port
F
IGURE
9-2:
I
NSTRUMENT
C
ONNECTIVITY FOR
EX1200A
DMM
C
ALIBRATION
Once the standards have been connected, the following connections should be made to the
EX1200A:
1)
Connect the EX1200A LAN port to the Ethernet switch using an Ethernet cable.
Connect the function generator, external DMM, and calibrator to the DMM as shown in Figure 9-3
using the EX1200A DMM calibration/verification cable (VTI P/N: 52-0807-000).
External DMM
Input I
Input HI
Input LO
Sense HI
Sense LO
Function Generator
Output HI
Output LO
Sync HI
Sync LO
Calibrator
Output I
Output HI
Output LO
Sense HI
Sense LO
EX1200 DMM
Input I
Input HI
Input LO
Sense HI
Sense LO
F
IGURE
9-4:
EX1200A
C
ONNECTIVITY TO
R
EFERENCE
S
TANDARDS
NOTE
Be certain that the calibrator is in standby mode (i.e. not outputting) before the cables are
connected to eliminate the possibility of having the function generator and calibrator generating
outputs at the same time while shorted together.
C
ONNECTING TO THE
EX1200A
There are three ways to connect to the EX1200A: using the IP address of the EX1200A, using
Bonjour for auto-discovery, or using the Agilent Connection Expert (ACE). For more information
on obtaining the EX1200A
’s IP address and using this as the conne
ction method, refer to the
. To connect to the EX1200A via Bonjour, refer to
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......