VTI Instruments Corp.
24
EX1200A Series Introduction
AC
F
ILTER
F
ILTER
T
YPE
B
ANDWIDTH
S
ELECTION
LOW*
3 Hz - 300 kHz
Can be selected only in 6½ Digits
MEDIUM*
20 Hz - 300 kHz
Can be selected only in 6½ Digits
HIGH
20 Hz - 300 kHz
Can be selected in 6½, 5½, and 4½ Digits
* Line frequency default to 60 Hz, integration time default to 0.01 PLC, and auto-zero settings default to ON in Low and Medium filter settings
AC
F
REQUENCY AND
P
ERIOD
*
R
ANGE
F
REQUENCY
R
ANGE
M
ODE
R
ESOLUTION
D
IGITS
1
Y
EAR
A
CCURACY
PPM
R
EADING
O
FFSET
PPM
100 mV to 300 V
3 Hz to 1.5 MHz
Slow
6
½
60
0.3
Medium
5
½
60
3.0
Fast
4
½
60
30.0
* Input signal > 10% of range.
T
EMPERATURE
A
CCURACY
-
T
HERMOCOUPLES
Type
Min
Max
-100
0
100
300
500
700
900
1100
1400
J
-200
1200
±0.25
±0.20
±0.20
±0.25
±0.30
±0.30
±0.35
±0.45
K*
-200
1372
±0.25
±0.20
±0.20
±0.20
±0.35
±0.35
±0.45
±0.55
±0.50
T**
-200
400
±0.25
±0.20
±0.20
±0.20
±0.25
E
-200
900
±0.25
±0.20
±0.20
±0.20
±0.25
±0.30
±0.35
S
-50
1768
±1.00
±0.75
±0.65
±0.65
±0.65
±0.70
±0.70
±0.75
R
-50
1768
±1.00
±0.75
±0.60
±0.60
±0.60
±0.60
±0.65
±0.70
B
250
1820
±1.65
±1.10
±0.80
±0.70
±0.65
±0.65
N
-200
1300
±0.40
±0.25
±0.25
±0.25
±0.30
±0.35
±0.40
±0.40
Values in °C
Conditions
•
Requires use of breakout box that is in a thermally stable environment.
• 30
minute warm-up
• Guaranteed maximum limits are two times (2x) the typical values
• Exclusive of thermocouple errors
• Exclusive of noise
• Common mode voltage = 0
* 1400 accuracy is for 1372 °C
** 500 accuracy is for 400 °C
ACV
&
ACI
R
EADING
R
ATE
F
ILTER
S
ETTING
R
EADING
R
ATE
LOW
3 readings/second;
Auto-zero ON, sample delay = 0, 0.01 PLC
MEDIUM
20 readings/second;
Auto-zero ON, sample delay = 0, 0.01 PLC
HIGH
2000 readings/second;
Auto-zero off, sample delay = 0, 0.01 PLC
S
YSTEM
S
PEED
E
VENT
Function Change
DCV & DCI: 9 / s
4-W R (100, 1K): 9 / s
2-W R (<1M): 1 / s
ACV: 0.4 / s, High Filter
ACI (0.01A & 0.1A): 0.25 / s, High Filter
Range Change
DCV & DCI: 300/s
4-W R (100, 1K): 300/s
2-W R (<1M): 2/s
ACV (0.1 to 100V): < 1.25/s, High Filter
ACI (0.01 to 0.1A): < 0.2/s, High Filter
Auto-Range Time
DCV & DCI: < 30 ms
4-W R (100 to 100K): < 60 ms
2-W R (100 to 100K): < 3.0 s
ACV (0.1 to 100V): < 2.0 s, High Filter
ACI (0.01 to 0.1A): < 4.0 s, High Filter
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......