www.vtiinstruments.com
EX1200A Series Maintenance
145
M
AINTENANCE
R
EPLACEABLE
F
USE
The EX1200A DMM has a replaceable fuse located on its front panel for current input and is user-
replaceable. It is a 3.15 A, 250 V fuse and can be replaced with VTI P/N: 35-0054-315.
Alternatively, P/N: 2173.15XP from Littlefuse, Inc. may also be used as a replacement.
CPU
B
OARD
R
EMOVAL
/R
EPLACEMENT
(3U
M
AINFRAMES
O
NLY
)
In the event that the CPU board of the EX1207A, EX1208A, and EX1209A mainframes requires
replacement, it can be removed and returned to the factory for repair or replacement.
Part Number
Description
Compatible Mainframe
70-0460-000
Final Assy, EX1208A CEL-4 CPU Board
EX1208A
The following sections describe how to remove the CPU board from the 3U mainframes.
NOTE
As these devices are static sensitive, proper ESD precautions should be exercised when removing
and handling the DMM.
CEL-4 CPU Board Removal
Removal Procedure
1)
Using the #2 Phillips screwdriver, loosen the jack-out screws from the CPU board front panel.
These screws are designed to pull the card out of the backplane’s mating connector. It is
recommended that both screws be loosed in equal amounts to avoid tension that may occur if
one screw is completely removed before the other.
2)
Using the jack-out screws, remove the CPU board from the mainframe.
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......