If a PTP master clock does not exist on the network, an LXI device on the network will begin
serving as the PTP time master. The
Time Configuration
page allows for PTP time to be set
manually. From the
Time Configuration
page, select the
Set Time
checkbox, then select the
Grandmaster
radio button from the
Target
field, and then configure the time by hand. Note that
the
Time
field should be completed using a 24-hour format.
F
IGURE
7-16:
PTP
C
ONFIGURATION
P
AGE
To use the EX1200A as the PTP clock that will serve time on the LXI network, select the
Self
radio button in the
Target
field. The EX1200A then becomes the PTP master clock. While this
ensures measurement synchronicity between the units that to this clock, the overall timestamp
accuracy is not as accurate because of the
time’s
hand-set nature. For higher overall timestamp
accuracy, use a PTP master clock that is synchronized to a global time source like, such as GPS.
Using SNTP/NTP Time
By selecting
SNTP
from the
Time Source
menu, the EX1200A can be synchronized with an
SNTP server. The
Time
Zone
drop menu allows the user to select the local time zone. SNTP, or
NTP (Network Time Protocol), servers are specified in the
NTP Servers
field, by IP address or
hostname, as shown in Figure 7-17. Once configured, the user must click
Submit
in order for the
changes to take effect.
To configure the EX1200A to use NTP time as its clock source:
1)
Select
SNTP
from the
Time Source
field. The page will change so that it can be configured.
2)
Add
/
Remove
NTP servers use the appropriate buttons until the EX1200A is configured to
communicate to the desired server(s).
3)
Click the
Save
button to save the server configuration.
4)
Click
Submit
to begin using NTP time.
NOTES
1) If the
Submit
button is clicked before saving an NTP configuration, the EX1200A will start
using NTP time, but will use the last previously saved server configuration.
2) Specifying SNTP or NTP servers by hostname requires that the instrument has DNS, either by
DHCP or Static IP.
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......