www.vtiinstruments.com
EX1200A Web Page Operation
137
NOTES
1)
The instrument password uses lower-case letters.
2)
This password mechanism provides only the most basic security. Applications requiring
additional security should consider using private, dedicated subnets, firewalls, etc. to
limit access to the instrument on the network.
U
PGRADE
This action page is used to upgrade the embedded firmware of the EX1200A controller and DMM
(if installed) as well as the plug-in modules. Prior to initiating the firmware upgrade process, a
new, uncompressed firmware image must be obtained from VTI Instruments and be accessible
from the computer that is connected to the EX1200A. Unless specifically noted by VTI
Instruments, firmware upgrades do not alter the non-volatile configuration settings (network
configuration, time configuration) of the EX1200A.
NOTE
Do not power cycle the EX1200A during the firmware upgrade process. If power is lost during the
upgrade, the instrument may be put into an inoperable state, requiring return to the factory. An
uninterruptible power supply may be used to avoid this risk.
F
IGURE
7-22:
U
PGRADE
W
EB
P
AGE
Perform the following steps to conduct a firmware upgrade:
1)
Remove Temporary Internet Files.
2)
Perform a
Reboot
or a power cycle.
3)
Connect to the EX1200A via the embedded web page.
4)
Click on the
Upgrade
link.
5)
Click on the
Browse
button and select the firmware image file to be uploaded to the
instrument.
6)
Click the
Submit
button to initiate the upgrade process.
During the firmware upgrade, an
Upgrade Status
window will appear (see Figure 7-23). Note
that, if a DMM is not installed in the system, one fewer step be reported. Controller/DMM
firmware upgrades typically take about 5 minutes to complete. Upon completion of the upgrade,
click on the
Reboot
button. After rebooting has completed (approximately one minute), click on
the
Index
link in the
Command
Menu
to continue operating the EX1200A using the web
interface. Note that the “Rebooting” prompt will not refresh itself.
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......