VTI Instruments Corp.
86
EX1200A Triggering
When using switch multiplexer modules such as the EX1200-3048, each input channel pair (HI
and LO) is routed through relays which are close to one another on the PCB to maintain the same
average temperature. Because of this proximity, the relays generate approximately equal
thermoelectric voltages on both the HI and LO channels, making the differential reading across
these two switches approximately cancel the thermoelectric voltages.
DMM
V
T1
V
T2
HI
LO
Switch Pair
V
M
V
S
= Source Voltage
V
T1/
V
T2
= Thermoelectric Voltage for Switch 1 and 2
V
READING
= Voltage Displayed by DMM
V
READING
= V
S
+ (V
T1
- V
T2
)
F
IGURE
4-4:
N
EGATION OF
S
WITCH
T
HERMOELECTRIC
V
OLTAGES
Measuring Voltages with High Output Resistance
Source loading errors occur when the output resistance of the DUT is a significant percentage of
the DMM’s
own input resistance.
R
S
V
S
HI
DMM
LO
R
i
V
S
= Ideal DUT Voltage
R
S
= DUT Source Resistance
R
i
= Multimeter Input Resistance
(10 M
or 10 G
)
Error % = 100 x R
S
R
S
+ R
i
F
IGURE
4-5:
E
RROR
C
ALCULATION FOR
H
IGH
O
UTPUT
R
ESISTANCE
L
OADS
To reduce the effects of source loading error and to minimize noise pickup, the input resistance of
the DMM can be set to HIZ mode (> 10
GΩ) for the 100
mV dc, 1 V dc, and 10 V dc ranges. The
input resistance is maintained at 10
MΩ for the 100
V dc and 300 V dc ranges.
Common Mode Rejection
An ideal DMM would be floating and isolated from earth-referenced circuits. In practice, though,
there is a finite resistance between the DMM
’
s input LO terminal and earth/chassis ground (see
Figure 4-6). This can cause errors when small voltages which are actually floating relative to earth
ground are measured (as for a photovoltaic cell, for example).
V
F
= Float Voltage
R
s
= DUT Source Resistance Imbalance
R
i
= Multimeter Isolation Resistance
Error (v) = V
f
x R
s
R
s
+ R
i
V
TEST
V
f
R
S
HI
LO
DMM
R
i
> 10 G
F
IGURE
4-6:
C
OMMON
M
ODE
R
EJECTION
E
RROR
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......