VTI Instruments Corp.
68
EX1200A Series Basic Mainframe Operation
F
IGURE
3-1:
T
RIGGER
B
US
C
ABLING
E
XAMPLE
As a general purpose output device, each Trigger Bus channel can be independently programmed
with regards to its output functionality and its static level to assume when enabled as an output.
When not enabled as an output, a channel becomes tri-stated, preventing conflict with other
potential voltage drivers. Regardless of output functionality, each channel provides constant input
functionality. That is, the input level on each channel can be accessed without a specific enable
function call. Moreover, the base functionality of the Trigger Bus channels is not affected by
triggering, scanning, or any other instrument process.
When enabled as an output, each channel also has the ability to generate a 1 µs pulse upon
command. The specific operation of the pulse depends on the static level programmed for that
channel. When a channel is programmed with a static level of high, the pulse will be low-going.
When a channel is programmed with a static level of low, the pulse will be high-going. Each pulse
generation requires a separate function call.
A common application of the pulse generation is multiple-unit synchronization. Referring to
Figure 3-1 above, each unit can be configured to trigger on the same Trigger Bus channel. Then
one unit is selected as a master and is used to output a signal on the designated Trigger Bus
channel line, serving as a common trigger source for all three instruments.
The default selections for each Trigger Bus channel are:
output enable is off
output level is 0
The electrical specifications for the trigger bus are provided in Table 3-4.
Characteristic
Value
Logic Type
M-LVDS Type 2
V
IT+
150 mV max
V
IT-
50 mV min
V
OS
1 V typical
T
ABLE
3-4:
T
RIGGER
B
US
E
LECTRICAL
S
PECIFICATIONS
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......