VTI Instruments Corp.
90
EX1200A Triggering
Waveform
Example
Crest Factor
Sine
1.414
Full-wave Rectified Sine
1.414
Half-wave Rectified Sine
2
Triangle
1.732
Square Wave (Symmetrical)
1
Repeating Pulse (PWM)
T
t
1
1
t
T
F
IGURE
4-10:
C
REST
F
ACTORS FOR
C
OMMON
W
AVEFORMS
Crest factor related errors are specified in the DMM specifications on page
Error! Bookmark not
defined.
. The specifications are for non-sine waveforms that are above 100 Hz when the Low
filter (3 Hz) is used. The measurement error can be determined for non-sinusoidal input using the
deratings specified in the datasheet.
Loading Errors
In ACV mode, the input of the EX1200A DMM appears as a 1
MΩ resistance in parallel with
100 pF of capacitance. The cabling used to connect the signals and the switch module used to
route the signals to the multimeter also add additional capacitance and loading. When connected to
high output impedance sources, like high-voltage probes, an error will be present as the result of
loading on the source. This error can be calculated using the following equation:
1
1
1
1
1
2
1
1
100
%
2
s
s
s
in
R
M
M
R
M
R
M
fC
Error
Where:
R
s
= source resistance
f
= input frequency
C
in
= input capacitance (100 pF) plus cable capacitance plus switch capacitance
Summary of Contents for Ametek EX1200A Series
Page 33: ......
Page 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Page 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Page 65: ......
Page 73: ......
Page 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Page 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Page 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Page 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Page 155: ......
Page 161: ......