03-914642-00:6
2-1
2
Site Preparation / Installation
Please read this section carefully before you begin to assemble the Archon.
2.1
Materials Needed
Regulator, single, two-stage with stainless steel diaphragm
In line on/off valve
Gas supply line tubing, minimum 1/8" diameter
Helium gas, 99.999% purity
Tools, 1/4", 1/16" open end wrenches; 3/16" blade screwdriver, Phillips screwdriver
Remote I/O Cable for communication to concentrator, see Section 8.
2.2
Unpacking the Archon
Each Archon is shipped in two cartons. The sample tray, accessory kit and
installation kit, with any optional spare parts, are shipped in a separate carton
from the Archon.
WA R N IN G
Avoid back strain or injury by following all safety precautions
when lifting heavy objects.
Carefully open the shipping cartons. Lifting from the bottom remove the
Archon from its carton (the largest box). Place it on the bench to the left side
of the concentrator.
Remove the sample tray, interface kit, interface cable, and any accessory or
spare items from the second carton.
Inspect for possible shipping damage or shortages. If damage is discovered,
immediately notify the carrier and Varian.
Varian Analytical Instruments
2700 Mitchell Drive
Walnut Creek, California 94598-1675
Attention: Manager of Customer Service
925-939-2400
Outside the U.S.A., notify the nearest Varian office.
Store the shipping cartons and all packing material for possible future use.
Inspect the contents against the packing list to determine if all items ordered
were received.
Locate the shipping, locking screw with spacer, on the left side of the cabinet.
This screw and spacer must be removed prior to turning the Archon on. Use a
small screwdriver to remove the screw. Save the screw and spacer for
possible future use.