System Options
The *System Options* menu allows for enabling optional features of the Archon.
When *System Options* is selected, the following menu is displayed. The “2
”
and “8
” keys are used to scroll through the menu lines. The “4
” and “6
”
keys are used to toggle the “NO - YES" response. The ENTER key will accept the
selection. Note: The following menu shows the system default settings as
shipped from the factory.
* System Options *
<Barcode Scanner?
NO
>
Needle Sparge?
YES
Ign. Vial Type
NO
Ignore No Vial?
YES
Hot Water Rinse?
YES
Vial Checks?
YES
Beep on Error
YES
Barcode Scanner:
If the Archon is equipped with an optional Barcode
Scanner, answering “YES” will tell the Archon Robotic
Arm to place the sample vials in the Barcode Scanner
Zone prior to vial type detection. The barcode label on the
vial will be read and the information may be retrieved to a
PC for its use. Only one barcode label will be retained for
each vial position.
Needle Sparge:
If the Archon is setup for Needle Sparge sampling, this
option must be “YES.” The sample vial type sensor will
then be reversed to detect water (40 ml VOC) vials. If a
SoilVial is detected, an error will be reported.
Ignore Vial Type:
This option disables the sample vial sensor. When “YES"
is selected, only the “sample type" parameter will
determine the sampling location. See “Method Setup” for
further details.
C A U T IO N
If “YES is selected, SEVERE damage will occur if the
sample tray is not loaded correctly. SoilVials and Water
vials MUST be installed per the Sample Type parameter.
See Section 5.2 Method Parameters for more details.
See Section 1.5 for vial types