System Setup
03-914642-00:6
6-17
Ignore No Vial:
When this option is “YES,” tray positions that have no
vials will be ignored. The Gripper “No Vial” sensor will
always be used to detect the presence of a vial and the
potential of lost vials in transit. The default should be
“NO”.
Hot Water Rinse:
When this option is “YES”, the “Hot Water" reservoir is
heated during the syringe and sparge tube rinses. This
reservoir is located inside the rear or the unit. Water
passes through the heated reservoir prior to entering the
25 ml syringe and sample transfer lines.
Vial Checks:
This option will cause the Robotic Arm to search the
equilibrium and vial sampling locations for vials during
start-up. The Archon will remove the vials and return
them to the sample tray.
C A U T IO N
This option should be set to “YES” to prevent vial jams.
Beep on Error:
This option causes the Archon to produce an audible
“Beep” when a System error occurs. The beep will sound
every five seconds. This would be anything which may
interrupt or stop the sample run. If a problem occurs and
the Archon “Beeps”, immediate action must be taken to
resolve the problem. An error message should be
displayed indicating the problem.