Site Preparation / Installation
03-914642-00:6
2-9
this does not allow the Archon to be installed. Therefore, the 3-port fitting must be
replaced with a 4-port cross fitting.
1. Remove the top, left side and valve oven covers from the Tekmar 3000.
2. Using a screwdriver, loosen the screw which secures the thermocouple and
the clamp on the three-port fitting to the cabinet.
3. Using a 1/4" open end wrench, loosen the nuts in all three ports.
4. Loosen the two screws, which secure the sparge tube mount to the cabinet.
5. Pull the sparge tube mount and tubing, out of the 3- port fitting and off the
instrument.
6. Pull each of the other two lines out.
7. Replace the original fitting with the new 4-port cross fitting.
8. Attach all of the lines into the new fitting. Replace the sparge tube, with
mount, back on the cabinet.
Refer to Figure 2-6 for correct line placement in the new fitting.