Site Preparation / Installation
03-914642-00:6
2-5
Figure 2-2
Insert the cap, with the PEEK pickup line, into the bottle, screw the cap onto the
bottle.
NOTE: DO NOT OVERTIGHTEN THE CAP. SLOWLY TURN THE CAP
ONLY UNTIL SNUG. OVERTIGHTENING WILL SPILT THE CAP.
Verify the helium is turned on at the source. Turn the toggle valve for blank water
reservoir and standard helium to the ON position. The valve is located inside the
cabinet on the right rear wall, see Figure 2-3.
To test for a pressurized bottle, depress the vent button located on the blank
waste bottle plug. You should hear the release of pressure. To refill the bottle,
turn the toggle switch OFF and vent the bottle by depressing the chrome button.
When the pressure is completely vented, remove the cap. Fill the bottle as
described above.