9-2
Message
Description
Lost vial.
Press any key to reset.
This indicates the gripper has lost a vial. Check for a
jammed vial or damaged sensor.
Replace vial (#) in the tray.
Press any key to continue.
This message indicates a vial was removed from the
tray or Equilibrium Block. Examine the vial locations
and replace the vial.
Could not step off vial.
Press any key to reset.
This indicates the vial sensor switch is continuously
detecting no vial. This could be due to no helium, a
broken switch or a mechanical problem with the
gripper.
Stuck vial.
Press any key to reset.
This indicates the vial is jammed or stuck in the
gripper. This could be a damaged vial sensor switch.
Could not sense vial(#) in vial
type slot.
Press any key to reset.
This message indicates the vial located in the vial ID
(type) cup was not detected. This could be a damaged
vial sensor switch, incorrect coordinate, or a
mechanical problem.
Soil vial can't be run as
Needle Sparge.
Press any key to continue.
This indicates a SoilVial was identified in the ID cup
and the System Option Needle Sparge is YES. See
Section 6.7. An adjustment may need to be made on
the photo optic sensor, or the sensor is bad.
Water/Soil vial can't be run in
a Water/Soil method.
This indicates the vial type doesn't match the method
type.
Remove vial (#) from the tray
and then
Press any key to continue.
This indicates the gripper is trying to load a vial in a
position which already contains a vial. Remove the vial
from the position and continue.
Didn't sense vial. Put it
in the tray and press any key.
This message indicates the gripper tried to move a vial
back to the tray, and no vial was detected. Check the
Equilibrium Block and tray for any vials which may be
loaded incorrectly.
(Elevator or Syringe) already
homed error. Pos (#).
Press any key to reset.
This indicates the Elevator or Syringe motor arrived too
early to its home position. This could be a slipping belt,
motor or electronic problem. (Elevator or Syringe) not
homed error.
(Elevator or Syringe) not homed
error.
Press any key to reset.
This message indicates a motor has not reached it's
home position. This could be a slipped belt, a failed
motor, board or home switch adjustment.
(Elevator or Syringe) home
error.
Press any key to reset.
This is only on start up. It means the Elevator or
Syringe motor never homed. The motor is for some
reason out of position. This could be a bad belt, motor,
switch or electronic problem.
(Right/Left) get position
timeout.
Front/Rear
Vertical
Press any key to reset.
This is a communication failure between the motor
board and processor board. It could be due to a poor
connection or defective cable.