Cleaning Procedures
The cleanliness of the sources, ion trap and conversion dynode can have a
significant impact on the performance of the mass spectrometer. The frequency
of cleaning depends on the quantity and nature of the samples run, so no
standard cleaning interval can be recommended. The troubleshooting guide in
this manual describes some of the symptoms that result from dirty components
and makes recommendations about when to perform cleaning procedures.
Cleaning the External Source
WARNING:
SHOCK HAZARD
Dangerous high voltages are present. Unplug instrument power cord.
•
Go to “Turning Off the Mass Spectrometer” on page 34.
•
Go to “Removing the Analyzer Assembly” on page 37.
•
Go to “Removing the Source/Ion Trap Assembly” on page 41.
Removing the Source Holder
1. Remove the two screws holding the source to the magnet structure.
2. Lift off the source assembly and place on its side on a lint-free cloth.