NEMA 5-20P, NEMA 5-15P, and CEE 7/7 Power Plugs and Outlets
With a 120V power source, the maximum amperage requirements for each of the
4000 MS components are as follows:
Component Amperes
Mass Spectrometer
12
Gas Chromatograph
20
Varian 8400 AutoSampler
0.5
Computer
3
Monitor
2
Laser Printer
3-4
NOTE: With a 230V power source, the maximum amperage requirement of each
of the above components is one-half of the amperage given above.
Never plug the mass spectrometer and the gas chromatograph into the same
power source; otherwise, you may overload the fourplex power source. The
Interconnect Diagram for the 4000 MS shows the five power cables of a typical
installation. Never use the free outlet on each of the power sources for
equipment that draws more than 2A.