Site Requirements
Site Preparation
The 4000 MS has been designed to operate reliably under carefully controlled
environmental conditions. It is the responsibility of the purchaser to provide a
suitable location, a power source of acceptable quality, and a suitable operating
environment. Operating a system or maintaining it in operational condition
outside of the power and operating environment limits listed below could cause
failures of many types. The repair of such failures is specifically excluded from
the standard warranty and service contract coverage.
For additional information, please request specific pre-installation support directly
through your local Varian Sales/Service Center.
Power
You are responsible for providing two dedicated fourplex single-phase power
sources with earth grounds hard-wired to the main power panel ground. Within
North America or Japan these power sources must be 20A, 90-130 Vac, 60 Hz
±
3 Hz, and outside North America they must be 10A, 180-260 Vac, 50 Hz
±
3 Hz.
One of these fourplex power sources is for the mass spectrometer, computer,
monitor, and printer. The other fourplex power source is for the gas
chromatograph and (optional) autosampler. If you have additional sample