4. Press out on the release tabs to remove the controller to manifold cable (1).
5. Pull on the white pull-tab to remove the manifold lens cable (2).
6. Press down on the locking connector and pull out to remove the manifold
power cable (3).
Verify that the transfer line is fully retracted and locked into position to prevent
damaging its tip.
7. Lift the analyzer assembly up and out of the manifold and place it upside
down on the work area.
8. Cover the manifold opening with a Kimwipe® or other low lint material to
avoid dust.
9. From this point, use clean tools and wear powder-free gloves.