6.FUNCTIONS
62
6-8. Digital Filter
This function minimizes instability of the weight value through moving average of A/D-
converted data. The number of times of moving average can be selected in the range of 2 - 256.
The larger the number of times of moving average, the more the weight value becomes stable,
but the slower the response becomes. On the contrary, the smaller the number of times of
moving average, the quicker the response becomes, but the more the weight value becomes
unstable. Select the most appropriate value according to the type of weighing.
6 : 64times
4 : 16times
8 : 256times
5 : 32times
Setting /
2 : 4times
0 : OFF
7 : 128times
3 : 8times
1 : 2times
CC-Link
0
30
MSB
LSB
RWw0006
RWw0005
RWw0004
Write the code No. and set value.
H
L
0
0 to 8
Code No.
Set value
remote register
Word size
Console
1) Select code 30.
2) Press the ENT key, and then set the
digital filter.
3) Determine the setting with the ENT key.
Digital filter