4.CALIBRATION
46
Resistance
Strain
Calculated value
approx. value
μ
-STRAIN
m
V/V
875 k
Ω
866 k
Ω
200
0.1
437 k
Ω
442 k
Ω
400
0.2
291 k
Ω
294 k
Ω
600
0.3
219 k
Ω
221 k
Ω
800
0.4
175 k
Ω
174 k
Ω
1000
0.5
146 k
Ω
147 k
Ω
1200
0.6
125 k
Ω
124 k
Ω
1400
0.7
109 k
Ω
110 k
Ω
1600
0.8
97 k
Ω
97.6 k
Ω
1800
0.9
87.3 k
Ω
86.6 k
Ω
2000
1.0
79.4 k
Ω
78.7 k
Ω
2200
1.1
72.7 k
Ω
73.2 k
Ω
2400
1.2
67.1 k
Ω
66.5 k
Ω
2600
1.3
62.3 k
Ω
61.9 k
Ω
2800
1.4
58.2 k
Ω
57.6 k
Ω
3000
1.5
54.5 k
Ω
54.9 k
Ω
3200
1.6
51.3 k
Ω
51.1 k
Ω
3400
1.7
48.4 k
Ω
48.7 k
Ω
3600
1.8
45.9 k
Ω
46.4 k
Ω
3800
1.9
43.6 k
Ω
43.2 k
Ω
4000
2.0
41.5 k
Ω
41.2 k
Ω
4200
2.1
39.6 k
Ω
39.2 k
Ω
4400
2.2
37.9 k
Ω
38.3 k
Ω
4600
2.3
36.3 k
Ω
36.5 k
Ω
4800
2.4
34.8 k
Ω
34.8 k
Ω
5000
2.5
*For reference, the relationship between the values of resistance connected
・
This table is for a 350 ohm load cell.
・
The temperature coefficient of the connected resistor directly influences
the accuracy of the indicator. Use a resistor with a temperature coefficient
below 50 ppm/
℃
. (approximately 5ppm/
℃
is recommended)
and input signals is shown in the table below.
Summary of Contents for F156
Page 1: ...F156 WEIGHING CONTROLLER 07 JUN 2012 REV 1 40 OPERATION MANUAL CC Link Ver 1 10...
Page 115: ...7 REMOTE CONSOLE UNIT 106 Monitor change Change the display with the key...
Page 120: ...7 REMOTE CONSOLE UNIT 111 Monitor change Change the display with the keys 0 to 9 1 2 3 4 5 0...
Page 155: ...13 DIMENSIONS 146 13 DIMENSIONS...