2.CONNECTION
7
CAUTION
・
The load cell excitation voltage of the F156 is 10V. Unless its maximum
excitation voltage is 10V or more, the load cell may generate heat or be
damaged.
・
For connecting a 4-wire load cell to the F156, be sure to cEXC and
+S, and –EXC and –S. Even if +S and –S are not connected, normal
operation will seem to be performed, but the load cell may generate heat or
be damaged because excessive voltage is applied.
+IN
-OUT
-IN
+OUT
Load cell
1
2
3
4
5
6
7
+EXC
+S
-S
-EXC
+SIG
-SIG
FG
Be sure to
connect here.