6.FUNCTIONS
92
●
The complete signal output timing depends on the setting of the complete output mode of
setting code 26.
●
Over/under comparisons are made, holding the weight value, when the complete output is
ON, deregarding the setting of over/under comparison mode of setting code 27.
●
Upper/lower limit comparisons are always made deregarding the setting of the upper/lower
limit comparison 2 of setting code 28.
t1 : Comparison inhibit time; comparison inhibit time of setting code 10
t2 : Judging time; judging time of setting code 11
t3 : Complete output time; complete output time of setting code 12
◆
Conditional expressions
・
Near zero:
ON when the weight value
≦
near zero set value.
・
SP1 output:
OFF when the weight value
≧
final set value – set point 1 set value.
・
SP2 output:
OFF when the weight value
≧
final set value – set point 2 set value.
・
SP3 output:
OFF
when the weight value
≧
final set value – compensation set value.
* These turn ON at start-time.
・
Under:
ON when the weight value < final set value – under set value.
・
Over:
ON when the weight value > final set value + over set value.
・
Go:
ON when the final set value + over set value
≧
weight value
≧
final set value – under set value.
●
Select the near zero comparison weight value from gross/net with setting code 22.
●
Select the SP1/SP2/SP3 output signal and over/go/under judgment signal comparing weight
value from gross/net with setting code 20.
Summary of Contents for F156
Page 1: ...F156 WEIGHING CONTROLLER 07 JUN 2012 REV 1 40 OPERATION MANUAL CC Link Ver 1 10...
Page 115: ...7 REMOTE CONSOLE UNIT 106 Monitor change Change the display with the key...
Page 120: ...7 REMOTE CONSOLE UNIT 111 Monitor change Change the display with the keys 0 to 9 1 2 3 4 5 0...
Page 155: ...13 DIMENSIONS 146 13 DIMENSIONS...