SECTION 8
TURBO MOLECULAR PUMP RECONDITION
62
8
8.1
Recommended Maintenance Intervals
It is different for deterioration progress speed of each part changes greatly by pump condition.
Refer to the following list as overhaul of each process.
These are not terms of warranty.
(1) Recommended maintenance intervals for parts
These are not terms of warranty.
1) Non-active gas and Light load process
(2) Recommended maintenance intervals for other parts
1) It has the possibility that the following part is exchanged, when cleaning TMP.
2) Others
When exchanging parts, it has possibility that other parts are exchanged for improving a
reliability.
(Example : Old type lower and upper radial sensor had changed dimensions for reliability.
When exchanging them, these housing must also be exchanged.)
Process
Recommended
maintenance intervals
1
Non-active gas (Sputtering, Evaporation and so
on) and Light load process
3 years
Part name
Recommended
maintenance intervals
1
Touch down bearing
3 years
2
Shaft
7 years
It has a possibility of under 2 years
(Dependent on condition)
3
Rotor
4
Motor
7 years
5
Magnetic bearing parts
Part name
55-pin Receptacle
Summary of Contents for UTM3400 Series
Page 2: ...This page is intentionally left blank...
Page 10: ...viii Location of Label...
Page 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Page 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Page 30: ...16 3 3 3 Controller This page is intentionally left blank...
Page 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Page 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Page 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Page 73: ...7 7GAS PURGE...
Page 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Page 95: ...10 10WARRANTY CLAUSES...
Page 98: ...84 10 This page is intentionally left blank...
Page 100: ...Index Index 2 This page is intentionally left blank...