SECTION 6
OPERATION
44
6
6.4.2 S
hutting Down Sequence in LOCAL Mode
(1) Maintained push the START/STOP button (Fig. 2-2 (3)) and check that the STATUS green
lamp (Fig. 2-2 (7)) is turned off.
(2) Wait until the ROTATION lamp (Fig. 2-2 (6)) goes out.
(3) Turn off the POWER switch.
(4) Stop the cooling water flow.
(5) If you want to isolate this product from electrical supply, unplug AC INPUT connector.
6.4.3 S
hutting Down Sequence in REMOTE-control Connector
(1) Input the "STOP" signal (Refer to Table 6-4) from the remote-control connector.
If stop from "NORMAL" status, "NORMAL" signal is OFF. In addition, and check that the
"BRAKE" signal (Refer to Table 6-4) is ON in this product of a standard spec.
(2) Wait until the "ROTATION" signal (Refer to Table 6-4) turns off. At this time, the "BRAKE"
signal (Refer to Table 6-4) also turns off in the device of a standard spec.
(3) Turn off the POWER switch.
(4) Stop the cooling water flow.
(5) If you want to isolate this product from electrical supply, unplug AC INPUT connector.
6.4.4 S
hutting Down Sequence in Other Communication Means
(1) Input the "STOP" command via the communication means.
The pump deceleration starts. Status is changed from "NORMAL" or "ACCELERATION"
to "BRAKE".
(2) Wait until status changes to "STOP" status.
(3) Turn off the POWER switch.
(4) Stop the cooling water flow.
(5) If you want to isolate this product from electrical supply, unplug AC INPUT connector.
When the turbo molecular pump is turned off after pumping a corrosive gas, maintain a vacuum
inside the turbo molecular pump or purge the interior of the pump with an inert gas.
Further, in such a case when a hydraulic rotary vacuum pump is used as backing vacuum
pump and there is possible reverse flow and diffusion of oil from the backing vacuum pump,
return the pump internal pressure to atmospheric pressure using dry nitrogen gas, after complete
shut-down of the pump ROTATION lamp goes out, to prevent the turbo molecular pump from
being contaminated with oil vapor.
For shutting down the turbo molecular pump in running at high speed with infeed of dry nitrogen
gas to the pump, keep the nitrogen gas flow rate at 1500 mL/min maximum.
REFERENCE
The ROTATION lamp goes out, "ROTATION" signal (Refer to Table 6-4) turns off, or change
to stop status when the pump rotational speed is 60 rpm or less. Turning off the POWER switch
permits the pump rotor to be supported by the touch-down bearings.
Summary of Contents for UTM3400 Series
Page 2: ...This page is intentionally left blank...
Page 10: ...viii Location of Label...
Page 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Page 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Page 30: ...16 3 3 3 Controller This page is intentionally left blank...
Page 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Page 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Page 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Page 73: ...7 7GAS PURGE...
Page 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Page 95: ...10 10WARRANTY CLAUSES...
Page 98: ...84 10 This page is intentionally left blank...
Page 100: ...Index Index 2 This page is intentionally left blank...