3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 69
6. The ‘SHIFT’ key will illuminate to indicate that a change has been made
to the calibration of the instrument, however has not yet been stored.
7. To undo the adjustment before storing the changes, press the
UNDO
soft key
This will remove any changes that have been made to the output of the calibrator.
7. Measure and check the output again and then press
STORE
soft key
The following 2 screens are displayed briefly to confirm that the calibration factors
have been saved.
After displaying these messages, the shift key will also cease to be illuminated
SHIFT
SHIFT
Note: All the calibration points can be adjusted prior to storing the
calibration factors (
STORE
), however if the calibration routine is
ended or there is a power failure the new calibration factors will not
be saved if
STORE
has not been pressed.