Oscilloscope – Amplitude
1.
Select
CALIBRATE
using the soft key
2. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required parameter
‘Scope’ and then press
SELECT
soft key
3. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the 10mV/Div range and
then press
SELECT
soft key