3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 31
56. When calibration of this range is complete press the
RANGE
soft key
Select ‘AC Volts’ and then ‘2V AC’ and proceed with this range as described above.
57. The ’20V AC’, ‘200V AC’ and ‘1kV AC’ ranges are calibrated in the same
manner.
Note: The number of frequency (Hz) calibration points will
depend upon model of calibrator (3010A, 3041A or 3050A).