3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 36
56. When calibration of this range is complete press the
RANGE
soft key
Select ‘AC Amps’ and then ‘2mA AC’ and proceed with this range as described
above.
57. The ’2mA AC’, ‘20mA AC’,‘200mA AC’, ‘2A AC’ and ‘30A AC’ ranges are
calibrated in the same manner.
Note: The number of frequency (Hz) calibration points will depend upon model
of calibrator (3010A, 3041A or 3050A).
Remember to change the connections when calibrating the
30A range
Current connections - 200
A, 2mA, 20mA, 200mA
and 2A ranges
High current connections 30A range