3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 46
3. The Calibrator will now change to the 100 Ω Active Resistance range. Using the
Range up / Range Down keys, select the ‘Zero Calibration’ adjustment point
4. Connect the calibrator Voltage output terminals (Active Resistance output)
Ensure that the multimeter has been zeroed with the leads shorted. If applicable use
low current resistance (Low I) measurement modes. This is because active
resistance is a lower accuracy output typically used for calibration of 3 ½ and 4 ½
digit multimeters that use lower measurement currents.
5. Press and measure the output.
6. To adjust the output enter the value as measured on the multimeter
E.g. measured output = 29.95 Ω, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control
OUTPUT
ON
9
•
9
5
ENTER
2