3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 37
2Wire Ω
To calibrate the 2Wire Ω parameter, complete the procedure as follows:
NOTE : The 2Wire Ω output from the 3000A series calibrator is a passive output.
This means that the value on the display is the resistance value generated at the
terminals.
The resistance cannot be ‘altered’ to a different output, the value stored is simply the
value of the resistor and the connections to the terminal.
For a variable resistance output, the Simulated Resistance output must be used.
1.
Select
CALIBRATE
using the soft key
2. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required parameter ‘2
Wire Ω’ and then press
SELECT
soft key
3. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required range e.g.
’10 Ω ’ and then press
SELECT
soft key