3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 49
Capacitance
To calibrate the Capacitance parameter, complete the procedure as follows:
NOTE : The Capacitance output from the 3000A series calibrator is a passive
output. This means that the value on the display is the capacitance value generated
at the terminals.
The capacitance cannot be ‘altered’ to a different output, the value stored is simply
the value of the capacitor and the connections to the terminal.
1.
Select
CALIBRATE
using the soft key
2. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required parameter
‘Capacit.’ and then press
SELECT
soft key
3. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required range e.g.
’1uF ’ and then press
SELECT
soft key