3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 25
13.
Measure the output, and adjust as required using the process described in
steps
7 – 11
14.
Use the ‘range up’ and ‘range down’ keys to change the calibration point to
‘Negative Full Scale’.
15.
Measure the output, and adjust as required using the process described in
steps
7 – 11
16.
To continue adjusting other ranges in the DC Current function, select ‘DC
Amps’ and select the required range
17.
The ‘2mA DC’, ‘20mA DC’, ‘200mA DC’, ‘2A DC’ and ‘30A DC’ ranges are
calibrated in the same manner.