3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 83
The screen of the 3000A will now display 0V. If the reading is different, repeat the
previous stages
3. Using the Range Up / Range Down keys to select the Positive Full Scale
adjustment point
4. Apply an output of 10V from your voltage source to the adapter interface, using
Pin 7 as the Ground or Negative connection, and Pin 9 as the Signal or Positive
connection.
5. Wait for the reading to stabilise on the screen of the 3000A. After achieving a
stable reading, press the ‘Store’ button.