3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 70
8. Use the ‘range up’ and ‘range down’ keys to change the calibration point to
Scope BW 10MHz.
10.
Adjust as steps
ASJDASJDASDJ.
11.
Continue adjustment until 260MHZ (SCP250) / 360MHz (SCP350) / 620MHz
(SCP600) point has been adjusted
12.
Verify Bandwidth Output flatness by performing a sweep of the full frequency
range, ensuring that flatness is achieved across the full range of the
oscilloscope function.