3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 28
3. The Calibrator will now change to the 200mV AC range. Use the Range Up /
Range Down keys to select the 206Hz Positive Full Scale adjustment point
4. Connect the calibrator output voltage terminals to the precision multimeter. Ensure
that the multimeter is in AC Voltage on the appropriate range.
5. Press and measure the output.
6. To adjust the output enter the value as measured on the multimeter
E.g. measured output = 199.990 mV, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control
OUTPUT
ON
1
9
9
•
9
9
0
ENTER