3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 40
56. When calibration of this range is complete press the
RANGE
soft key,
Select ‘2 Wire Ω’, and then select the required range.
57. The ‘0 mΩ’, ‘100 mΩ,‘1Ω’, ‘10Ω’, ‘1 kΩ’, ’10 kΩ’, ‘100 kΩ’, ‘1 MΩ’, ’10 MΩ’, ‘100
MΩ’ and ‘1 GΩ’ ranges are calibrated in the same manner.